.. _yz_position: Sample Y and/or Z positions search ---------------------------------- There are 2 ways to search sample Y and/or Z positions: (1) Search by radiography image (2) Scan absorption profile (3) Scan by EDXD or flurescence profile (refer to :ref:`X position search `) By radiography image ^^^^^^^^^^^^^^^^^^^^ - Switch to :ref:`radiography measurement mode `. - Narrow slit size to those of EDXD measurement. - Mark the narrow slit size and position by tape or something on monitor. - Open slit (click :guilabel:`for Camera`) for radiography measurement. - Move sample position to x-ray beam position shown by a mark on monitor. By absorption scan ^^^^^^^^^^^^^^^^^^ - Move to :ref:`EDXD measurement mode `. - Confirm 2TH is >10°. - Open 'Bluediamond' (:ref:`Bluediamond ` software). - Select 'IC2 or Diode' in Detector in Bluediamond, uncheck all others. - Open 'scan' on SAM Y (or Z) .. figure:: /images/operation/sam_y_scan.png :alt: scan_launch :width: 720px :align: center - Set scan parameters of 'Start', 'End' and '#Pts' (#Pts has to be odd number) (confirm 'Relative'). .. figure:: /images/operation/sam_y_scan_window.png :alt: scan_go :width: 300px :align: center - Click :guilabel:`Load&Go` to start scan. - Scan results will appear in the ':ref:`Bluediamond `' window.